Job Card PHY A009 - Opposite Side Rotative Arm Installation
32. Route the mouse cable through the opening in the harness, so that it leaves the other end of the har-
ness. Then route the track cable through the rotative arm support entrance into the Control Station.
33. Ensure that the length of each of the monitor cables leaving the harness are at the distances as
specified below. Secure the harness to the monitor cables with a cable tie located at 1 cm from the
end of the harness.
34. Install and connect the monitor as described in
Job Card PHY A010 - Mount LCD Monitor on Control
-
.
Monitor video cable:
26 cm from harness end.
Monitor power cable:
26 cm from harness end.
Monitor earth cable:
6 cm from harness end.
26 cm
6 cm