GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A292 - Tube Tilt Pivots
Page no. 1399
Chapter 9
JC-DR-A-292.fm
7
COMPLETION
1. Switch ON the Gantry.
2. Switch ON the electrical power supply from the Mains Distribution Panel in the room.
Note:
Before performing the calibrations and checks listed below, carry out a detector clear calibration.
From the Service Desktop, click the
Calibrations
button, then select
Detector
and
Clear
Calibra-
tion
.
Calibrations / Checks
Job Cards
Acquisition Image Chain
Tube Tilt Calibration
Job Card CAL A027 - Tube Tilt Calibration
Collimator Calibration
Job Card CAL A029 - Collimator Format Calibration
Collimation Check
Job Card ELE A028 - Collimation Checks
If Collimation Check not OK, repeat Collimator Calibration
Bad Pixel Calibration
Job Card CAL A041 - Bad Pixel Calibration
Gain calibration
Job Card CAL A043 - Detector Gain Calibration
Flat field tests
Job Card ELE M032 – IQ Tools Flat Field Test on page 813
Mag Resolution Test
Job Card ELE M034 - IQ Tools Mag Resolution Test on page 835
If IQ Tests not OK, repeat all calibrations from Bad Pixel
Half Value Layer
Job Card CAL A046 - Half-Value Layer Measurement
IQ Tools Composite test
Job Card ELE M033 - IQ Tools Composite Test on page 821
If Composite Test not OK, repeat all calibrations from Half Value Layer
AOP: Automatic Optimization of Parameters
AOP Calibration
Job Card CAL A051 - AOP Calibration
mAs non linearity
Job Card CAL A053 - mAs Non-linearity Calibration
AOP/SNR Check
Job Card ELE A038 - AOP Mode and SNR Check
If AOP/SNR Test not OK, repeat all calibrations from AOP Calibration