GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 933
Chapter 9
ERR-SUB-A-003.fm
181/006
HALT
STEREOTAXY
HARDWARE FAIL-
URE
Hardware failure with the stereotix power.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the
Stereotaxy Service Manual
).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the
Stereotaxy Service Manual
).
181/007
HALT
STEREOTAXY
HARDWARE FAIL-
URE
Hardware engine failure with the stereotix A/D con-
verter.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the
Stereotaxy Service Manual
).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the
Stereotaxy Service Manual
).
181/008
HALT
STEREOTAXY NEE-
DLE PATH FAILURE
Security Failure: Needle path failure.
Try to reproduce the defect by playing with the nee-
dle path lock button.
• If defect is reproducible, the Z-axis trolley is faulty
and you must change the Stereotactic Positioner
(see the
Stereotaxy Service Manual
).
• If defect it is not reproducible, the faulty FRU is the
Stereotactic Positioner CPU board, and you must
change the CPU Board on the Stereotactic Posi-
tioner (see the
Stereotaxy Service Manual
).
181/009
HALT
STEREOTAXY
FAILURE
Security Failure: Power switch failure.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the
Stereotaxy Service Manual
).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the
Stereotaxy Service Manual
).
181/010
HALT
STEREOTAXY
SOFTWARE FAIL-
URE
Software failure with the stereotix sequencing.
1. Unplug and replug the stereotix then reboot the
system from the X-Ray Console.
2. If the problem persists, replace the stereotix unit.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card