GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A015 - Lift Movement Troubleshooting
Page no. 1047
Chapter 9
JC-TSG-A-015.fm
ILLUSTRATION 7 - TROUBLESHOOTING FLOWCHART FOR LIFT MOVEMENT PROBLEMS RELATED TO
ERRORS 102/013 AND 102/014 - TYPE 3 LIFT SCREW ASSEMBLY
Perform Lift Functional Checks.
See
,
.
NOK
Grease Checks
See
,
Grease Lift Screw
See
NOK
OK
NOK
Lift Gas Spring Checks
See
NOK
Change Lift Gas Spring
See
.
Perform Lift Functional Checks.
See
OK
NOK
Perform Lift Functional Checks.
See
,
.
OK
OK
Nut, Bearing, and Lift Checks
See
OK
If Nut, Bearing or Lift NOK,
change Type 3 Lift without Mo-
tor (5373048), keeping the mo-
tor from the old lift.
See
NOK
Perform Lift Functional Checks.
See
.
OK
Check Lift Motor
See
.
Note:
If you have just changed
the Lift motor as instructed
above in illustration 6, the Lift
Motor can be assumed to be
OK, so proceed directly to
changing the Lift Board.
NOK
Change Type 3 Lift Motor
(5413527)
See
.
Perform Lift Functional Checks.
See
,
OK
NOK
Contact OLC.
Corrective actions
Change Lift Board
See
NOK
Perform Lift Functional Checks.
See
,
OK
NOK
OK