Chapter 9
Page no. 1514
JC-DR-A-321_DS.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A321 - IDC Computer
16. Remove the six allen screws (9) that attach the baseplate to the IDC (2.5 mm allen wrench).
6-5
Reassembly of IDC Computer
Reverse the disassembly procedure.
See the table and illustration below for the IDC rear panel connections.
See
for the UPS rear panel connections.
9
9