Chapter 9
Page no. 884
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
122/021
HALT
COMPR FORCE
SENSOR DEFECT
Force has gone out of the acceptance range for the
compression force sensor.
Compression force out of range, refer to
A1217 COMPRESSION CONTROL FAILURE
This error can be due to a too high reverse force
applied on the paddle. If the reverse force is too sen-
sitive, the Faulty FRU is the Compression trolley kit.
A1217
122/022
N/A
EMERGENCY
STOP MOTION
Somebody presses the Emergency stop button.
Then the compression node cuts the stop motion line
and all nodes see it. Moreover, the compression
board sends this signal to Gantry CPU.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
CAL A048
CAL A050
122/024
DEBUG
ONLY
COMPR SOFT-
WARE ERROR
Occurs when the compression node is currently per-
forming a movement and the Gantry CPU sends a
command for another movement while the current
movement is not finished. This is a CPU problem
because the CPU did not detect the current move-
ment.
Ignore this error. If the problem persists, contact your
OLC.
122/025
HALT
COMPRESSION
NODE FW INCOM-
PATIBILITY
Occurs when the hardware configuration (CPLD) of
the node or the DSP or the MCU firmware not match.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
CAL A048
CAL A050
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card