Job Card DIAG A1000 - Gantry Diagnostic
FRUs:
2-33 JC DIAG A1215 COMPRESSION CONTROL FAILURE
Diagnostic for:
121/004 error code (CAN code is 0x9002) valid motion error
Error name: no valid line
Inspect the cables listed below. If you find any damage in a cable or connector, change it.
Check the cables in the following order:
-
W213 - POS-BUS 1 cable for Lift to Rotation;
-
W304A cable for paddle detector.
-
W304B cable for paddle detector.
-
W314 Sys Pos Bus cable.
-
W311 - POS-BUS 2 cable for Bucky board;
If the error persists, change the Compression Board (see
Job Card D/R A250 - Compression Board
).
If the error persists, change the Gantry CPU Board (see
Job Card D/R A190 - Gantry CPU Board
).
FRUs:
2-34 JC DIAG A1216 COMPRESSION COMMUNICATION FAILURE
Diagnostic for:
122/040 error code (CAN code is 0x8210) invalid PDO length
•
Change the Compression Board (see
Job Card D/R A250 - Compression Board
).
•
If the error persists, change the CPU Board (see
Job Card D/R A190 - Gantry CPU Board
). If changing the CPU board solved the problem, put back the original Compression Board (see
Job Card D/R A250 - Compression Board
).
FRUs:
2-35 JC DIAG A1217 COMPRESSION CONTROL FAILURE
Diagnostic for:
122/021 error code
(CAN code is 0x503D) force sensor range error
This error is reported when the value of the compression force is out of the -50% to +20% calibrated
Compression Board
Pot+cable
Compression Board
W314 Sys Pos Bus cable
W304A cable for paddle detector
W304B cable for paddle detector
W311 - POS-BUS 2 cable for Bucky board
W213 - POS-BUS 1 cable for Lift to Rotation
Nephtys Gantry CPU
Compression Board
Nephtys Gantry CPU