GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A010 - Set Lift Travel Limits
Page no. 569
Chapter 7
JC-ELE-A-010.fm
Potentiometer Upper Reference Voltage.
The Gantry LCD updates with the message
"Lift node fully calibrated. Calibration can be saved or
canceled on ADS."
and the AWS screen updates instructing you to either save or cancel the calibra-
tion (see
Selection of the Lift Travel Limit (Completion)
! Notice:
If you do not move the lift up to its highest possible position until it automatically stops, and you
press the FOV size button the Gantry LCD updates with the message
"Upper lift position out of
range. Check lift position or positioner signal! Then press FOV button."
. In this case simply contin-
ue to move the lift up to its highest possible position until it automatically stops and then press the
FOV size button.
6-2-4
Selection of the Lift Travel Limit (Completion)
Use the following steps to save the calibration results and set the lift travel limit,:
1. On the
Lift Position and Upper Limit Calibration
page, click the
Save
button.
A
Lift Optical Fork Position Calibration
page on the appears prompting you to select the upper optical
fork position.
2. Click the button corresponding to the upper optical fork position that you set in
-
click the
850 mm
button if the upper optical fork is in the upper position
-
click the
750 mm
button if the upper optical fork is in the middle position
-
click the
650 mm
button if the upper optical fork is in the lower position
-
click the
No Change
button if the upper optical fork has not been moved
3. From the
Lift Calibration SAVED
page that appears, click the
OK
button.
You are returned to the
Calibration: Positioner Summary
page.
4. From the
Calibration: Positioner Summary
page, click the
Exit
button to leave the Service Desktop.
6-2-5
Check the Lift Lower and Upper Travel Limits
Move the Arm to the lower position, then to the upper position and check whether the Arm stops at the
right positions.