GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A031 - AOP Configuration
Page no. 697
Chapter 7
JC-ELE-A-031.fm
7-4
Setting and Checking the AOP Configuration on the Generator
7-4-1
Setting the AOP Configuration on the Generator
Use the following steps to set the AOP configuration on the Generator.
1. Select
SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE
on the X-ray Console.
2. From the
SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE
menu, select the
DOSE
mode.
3. Select
TAB_A
to set the DOSE mode to use Classic table (TAB_A), or
TAB_B
to set the DOSE
mode to use Profile table (TAB_B).
4. From the
SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE
menu, select the
STD
mode.
5. Select
TAB_A
to set the STD mode to use Classic table (TAB_A), or
TAB_B
to set the STD mode to
use Profile table (TAB_B).
6. From the
SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE
menu, select the
CNT
mode.
7. Select
TAB_A
to set the CNT mode to use Classic table (TAB_A), or
TAB_B
to set the CNT mode to
use Profile table (TAB_B).
8. Press the
SETUP
button five times on the X-ray Console to reach the checksum menu.
9. Select
CKSUM/CKSUM
to perform a checksum.
10. Keep pressing the
SETUP
button until you exit the configuration menu on the X-ray Console.
7-4-2
Checking the Configuration Set on the Generator
Use the following steps to check the AOP Configuration that is set according to your requirements for
each mode on the Generator.
1. Select
SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE
on the X-ray Console.
2. From the
SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE
menu, select the mode that you want
to check (
DOSE
,
STD
,
CNT
).
3. Check the value displayed to the right selected mode.
•
If this value is TAB_A, the selected mode is configured with the Classic table.
•
If this value is TAB_B, the selected mode is configured with Profile table.
4. Keep pressing the
SETUP
button until you exit the configuration menu on the X-ray Console.
7-5
Setting and Checking the AOP Configuration on the ADS Computer
The AOP STD mode on the ADS Computer must be set to match the AOP STD mode on the Generator.
7-5-1
Setting the AOP Configuration on the ADS Computer
Use the following steps to set the AOP configuration on the ADS Computer.
1. From behind the Browser, launch a command window (see
Launching a Command Line Window
A command window appears with an ADS prompt.
2. Log in as root as follows:
a. At the ADS prompt enter:
su -
b. When requested to specify the root password, enter:
operator
3. Change to the
~sdc/senovision/scripts
directory, by entering:
cd ~sdc/senovision/scripts