Chapter 9
Page no. 1032
JC-TSG-A-012.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A012 - Use of DB9 Adaptors to Reduce Intermittent IDC or ADS Communication Errors
TABLE 2 - IDC DB9 CONNECTIONS
ILLUSTRATION 3 - IDC DB9 CABLE CONNECTIONS
DB9 cable connection reference in
illustration
Cable Label
1
IDC - Gantry Conditioner Link 2226612
2
Detector Power Supply 2226621
3
IDC - Gantry Serial Link RT 2240227 or 2246615
Dione4
1
2
3
1
2
3
Dione5