If you cannot move by hand, change the brake.
-
If you still cannot move by hand, check the compression mechanism again for mechanical defects
•
If you can move the timing belt by hand but the "step lost" error is still reported, change the compres-
sion motor and encoder. See
Job Card D/R A244 - Compression Motor with Encoder and Compres-
•
If the "step lost" error is still reported, change the Compression Board (see
FRUs:
Compression Board
Motor +Encoder with cable
Compression current-off brake