Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A402 - Generator CPU Battery
Page no. 1551
Chapter 9
JC-DR-A-402.fm
7
COMPLETION
In normal circumstances there is nothing to do. However, if the Generator configuration was lost in one
of the following situations, the Generator must be reconfigured:
•
the original battery was too low before you arrived
•
a problem occured during the backup procedure and you failed to solder the battery correctly
If Generator reconfiguration is necessary:
•
carry out the following calibrations:
•
For Senographe systems that use AWS V43.0 or later (on the SB150) or AWS 53.0 or later (on the
U20), if the Senographe system does not use the default AOP Classic (AOP A) tables configuration,
restore as necessary the AOP configuration (see
Job Card ELE A031 - AOP Configuration
Calibrations - Checks
Job Card
Acquisition Image Chain
Heating Current Calibration
Job Card CAL A001 - Heater Current Scale Factor
Bias Calibration
Job Card CAL A003 - X-Ray Tube Focal Bias Voltage Scale Factor
Tube mA Measurement
Job Card CAL A009 - X-Ray Tube mA Measurement
kV Scale Factor
Job Card CAL A011 - Calibration of kV Scale Factor
mAs X-ray Calibration
Job Card CAL A012 - X-Ray Tube Heater Current
Bad Pixel Calibration
Job Card CAL A041 - Bad Pixel Calibration
Gain calibration
Job Card CAL A043 - Detector Gain Calibration
flat field tests
Job Card ELE A032 – IQ Tools Flat Field Test
If IQ Tests not OK, repeat all calibrations from Bad Pixel.
Half Value Layer
Job Card CAL A046 - Half-Value Layer Measurement
Conversion Factor
Job Card CAL A045 - Conversion Factor Measurement
Use the same dosimeter as that used for CAL A046
AOP: Automatic Optimization of Parameters
AOP Calibration
Job Card CAL A051 - AOP Calibration
mAs non linearity
Job Card CAL A053 - mAs Non-linearity Calibration
AOP/SNR Check
Job Card ELE A038 - AOP Mode and SNR Check
If AOP/SNR Test not OK, repeat all calibrations from AOP Calibration.