GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A192 - Arm Distribution Board
Page no. 1145
Chapter 9
JC-DR-A-192.fm
below.
Arm Distribution board cable connections:
8. Switch on the Gantry electrical power.
9. Wait until the Gantry boot is complete, and check that no error has been reported.
10. Switch off the Gantry electrical power.
11. Reinstall the right column cover; refer to
Job Card PHY A044 - Remove/Reinstall Gantry Covers
Connector
Label
Description of cable
(usually identified by code on a yellow label)
Cable color
J1
J1/PL301-W314 (2345920-2)
Gray
J2
J2/PL301-W315 (2345921)
Gray
J3
J3/PL301-W503 (5143848)
Black
J4
J4/PL301-W310 (5139723)
Gray
J5
J5/PL301-W307 (2371179)
Gray
J6
J6/PL301-W308 (2372181)
Gray
J7
J7/PL301-W409 (2345934)
Gray
J8
J8/PL301-W401 (2345926)
Gray
J9
J9/PL301-W411 (2345919)
Gray
J10
J10/PL301-W410 (2345918)
Gray
J11
No cable attached
N/A
J1
J2
J5
J7
J6
J9
J8
J10
J3
J11
J4