background image

 

1 Manual Updated May 19, 2020 
 

MK5 Operator's Manual 

 

 

 

 

Содержание MK5

Страница 1: ...1 Manual Updated May 19 2020 MK5 Operator s Manual...

Страница 2: ...backlit display allow you to read measurements with ease in a variety of lighting conditions Its unique Capture Value function makes remembering values effortless It allows you to more easily capture...

Страница 3: ...3 Manual Updated May 19 2020 Diagram of MK5...

Страница 4: ...through Turn this knob counterclockwise to open the High Pressure Passthrough Do not exceed finger tight when closing the knob Do not use any external tools to increase torque applied to this or any k...

Страница 5: ...5 Manual Updated May 19 2020 Diagram of MK5 Continued...

Страница 6: ...oovy Place for the Hook to be stowed when not in use The orientation of the hook can be reversed facing left or right 19 Battery Compartment Screws Unscrew all four Battery Compartment Screws before r...

Страница 7: ...7 Manual Updated May 19 2020 MK5 Display...

Страница 8: ...automatically power off after 30 minutes of no use This conserves power and saves you batteries if you forget to turn MAKO off between uses F Calibration CAL Mode indicator Shows when the unit is in...

Страница 9: ...19 2020 MK5 Water Flow Diagram MK5 Water Flow Details In the diagram above arrows start and end on valves or the water source This is an accepted 5 valve water flow configuration as prescribed by the...

Страница 10: ...paper at each assembly Six values can be captured in memory and on the display allowing you to capture the working pressure as well as all test values on any backflow assembly The capture button allow...

Страница 11: ...g will be captured is shown by an active position cursor Position 4 in the example display above To capture a value press the Capture button To clear a value press and release Back button To clear all...

Страница 12: ...ield good results or when the pressure is too volatile to test The rate of change graph also coincides with the sensor location inside the Mako when held in the proper vertical testing orientation Thi...

Страница 13: ...will be darkened as in the picture above The faster the pressure is increasing the more arrows will be darkened Decreasing Differential Pressure When the pressure reading is decreasing the arrows on t...

Страница 14: ...Off Auto Power Off APO turns the Mako off after 30 minutes of non use To disable the Auto Power Off function Hold the Back button while powering on the Mako The Auto Power off indicator APO will no l...

Страница 15: ...or later installed Range PSID Accuracy 70 F PSID 0 to 15 0 1 15 to 100 1 of reading 100 to 150 5 of reading 150 to 175 10 of reading Note for USC FCCCHR Manual 10 Approval a minimum of 0 2 PSID accur...

Страница 16: ...SIDE the WARRANTY contact Arbiter Inc or one of our calibration distribution or service partners Reversing any hose direction or operating the Mako without hoses other than those provided by Arbiter o...

Страница 17: ...a warranty Opening the product or breaching the product case in any way except for opening the battery compartment to change batteries will void the warranty Damage to or failure to replace the batte...

Отзывы: