Chapter 9
Page no. 858
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
102/020
N/A
STOP LINE WAS
ACTIVATED BY
CPU
The main CPU detected an error and activated the
stop motion line or did not deactivated it at all.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
102/021
HALT
LOW VOLTAGE
ERROR ON LIFT
NODE
There is a low voltage error. If the voltage of the
power supply falls below 24 V the node sends this
error.
Change the Lift board then perform Lift Calibration.
D/R A201
CAL A031
102/022
HALT
LIFT READ TIME-
OUT ERROR
The MCU has not reacted on the interrupt request
(IRQ) coming from the CPLD before 15 ms.
Change the Lift board then perform Lift Calibration.
D/R A201
CAL A031
102/023
HALT
LIFT CPLD MEM-
ORY ERROR
There is a CPLD memory error. Test data from the
CPLD is periodically read to ensure the proper com-
munication. If this error does not match the node
sends this error.
Change the Lift board then perform Lift Calibration.
D/R A201
CAL A031
102/024
HALT
UNCONTROLLED
LIFT MOTION
DETECTED. STOP
LINE ACTIVATED
There has been uncontrolled motion. If the motion is
performed when the lift node is not supposed to drive
it, the lift node sends this error and activates the
MCU_STOP_MOTION line which leads to 0x500A
error.
1. Check the lift operation in low speed.
2. If the problem persists, grease the lift screw.
3. If the problem persists, refer to
to fix the
problem.
A1013
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card