ERR SUB A003 - Gantry Error Messages
Page no. 911
Chapter 9
ERR-SUB-A-003.fm
160/008
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to set an attribute.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/009
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to publish an event.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/011
HALT
INTERNAL SOFT-
WARE ERROR
An error code was received from a known node, but
it was not handled by the Gantry CPU in the actual
release.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/012
HALT
AUTO-REBOOT
MAX RETRY
ATTEMPTS
REACHED
The Gantry CPU failed to boot up on 3 consecutive
attempts.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/013 ADVISORY
AUTO-REBOOT
RECOVERY
STARTED
The Gantry CPU failed to boot up and is retrying the
operation.
If system boots successfully, no action is required.
If three reboot attempts are made and fail a 160/012
HALT error is raised.
161/001
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU received an unknown UIF request.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
161/008 ADVISORY
LCD: A07 Lift is not
operational
The Gantry CPU received a lift UIF request while the
lift node is not operational.
Operator error. No service recommended actions.
161/009 ADVISORY LCD: A18 Rotation is
not operational
The Gantry CPU received a rotation UIF request
while the rotation node is not operational.
Operator error. No service recommended actions.
161/010 ADVISORY
LCD: A03 Compres-
sion is not opera-
tional
The Gantry CPU received a compression UIF
request while the compression node is not opera-
tional.
Operator error. No service recommended actions.
161/011 ADVISORY
LCD: A10 Motion
stopped. Multiple
movements are
requested
The compression motion was stopped because the
Gantry CPU received an inconsistent UIF request.
Operator error. No service recommended actions.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card