Chapter 9
Page no. 856
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
102/002
HALT
LIFT SPEED
ERROR
When a coding wheel defect is detected through the
complemented hardware (HW) lines, the IPMS stops
and forbids the motion and sends an error message.
See
JC DIAG A1005 LIFT CONTROL FAILURE
A1005
102/04
HALT
LIFT LOWSPEED
The signal coming from motor encoder contains four
phases. A, A complementary, B, B complementary.
The Lift node checks two things: complemen
sequence of phases.
Check the lift operation in low speed and then refer
to
JC DIAG A1013 LIFT CONTROL FAILURE
A1013
102/005
HALT
UNCONTROLLED
LIFT MOTION
There has been a Correlation Error. In case signals
from absolute coder (potentiometer) and coding
wheel (motor encoder) are not correlated.
Refer to
JC DIAG A1008 LIFT CONTROL FAILURE
to fix the problem.
A1008
102/006
HALT
LIFT FOOTSWITCH
DEFECT
SwitchDefect [i] (for i = 1 - 4) occurs when the two
signals from the button do not arrive at the same
time on the CPLD. [i] indicates the nature of the but-
ton and the side right or left. Double Footswitch con-
cerns 2 error signals and quadruple footswitches
concerns 4 error signals:
- Lift up
- Lift down
(- Compression up)
(- Compression down)
Refer to
to fix the problem.
A1001
102/007
HALT
LIFT FOOTSWITCH
DEFECT
102/008
HALT
COMPRESSION
FOOTSWITCH
DEFECT
102/009
HALT
COMPRESSION
FOOTSWITCH
DEFECT
102/010
HALT
LIFT OVER UPPER
LIMIT
The IPMS detected that a motion has reached the
upper electrical stop limit. It stops and, forbids the
motion and sends an error message.
Refer to
JC DIAG A1008 LIFT CONTROL FAILURE
to fix the problem.
A1008
102/012
HALT
LIFT OVER LOWER
LIMIT
The IPMS detected that a motion has reached the
lower electrical stop limit. It stops, and forbids the
motion and sends an error message.
Refer to
JC DIAG A1008 LIFT CONTROL FAILURE
to fix the problem.
A1008
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card