GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 861
Chapter 9
ERR-SUB-A-003.fm
102/036
HALT
LIFT TX CAN BUF-
FER OVERRUN
Possible causes:
• There has been a transmit queue overrun. The
transmit queue between the MCU and the DSP is
full and the message dropped.
• The TP1 connector on the Lift board is closed.
Proceed as follows:
Check whether a jumper exists on the TP1 connector
on the Lift board. In this case you must remove the
jumper from the TP1 connector on the Lift board.
If there is no jumper on the TP1 connector on the Lift
board, the Lift board is faulty. Change the Lift board
then perform a Lift Calibration.
D/R A201
CAL A031
102/037
HALT
LIFT MAILBOX
OVERRUN
The mailbox is overrun. The mailbox queue used in
the inter-task communication in the MCU is full and
the mails are dropped.
The Lift board is faulty. You must change the Lift
board then perform a Lift Calibration.
D/R A201
CAL A031
102/038
HALT
CPU HEARTBEAT
NOT RECEIVED BY
LIFT NODE
There has been a heartbeat failure. The Gantry CPU
did not send periodic "heartbeat" CAN message.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
102/039
DEBUG
ONLY
HEARTBEAT DIS-
ABLED ON LIFT
NODE
The Gantry CPU tried to disable the "heartbeat"
check of the lift node.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
102/040
HALT
LIFT INVALID PDO
LENGTH
The lift node receives a PDO with an invalid DLC.
Refer to
JC DIAG A1110 ROTATION CONTROL
to fix the problem.
A1016
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card