Job Card D/R A318 - LCD Monitor
positioned to avoid a collision with the radiation screen.
8
COMPLETION
•
If Stereo or 2D Biopsy Optical Localizer option exists, calibrate the newly installed LCD Monitor
according to
Job Card ELE A043 - AWS Monitor Calibration
.
•
If you have upgraded from an 18" or 19" LCD Monitor to a 21" 3MP LCD Monitor, perform all of
Card ELE A008 - Configuring the 21-inch 3MP Monitor
so that:
-
the ADS software is correctly configured for the 21" 3 MP Monitor
-
the 21" 3 MP Monitor mode is configured to use DICOM mode
-
21" 3 MP Monitor buttons a locked
•
If you have changed an existing 21" 3 MP Monitor, perform
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
, so that:
-
the 21" 3 MP Monitor mode is configured to use DICOM mode
-
21" 3 MP Monitor buttons a locked
The LCD Monitor contains Mercury, therefore properly dispose of the LCD Monitor in accor-
dance to your local laws.