Job Card D/R A406 - Generator Command Board 400-PL1
Page no. 1575
Chapter 9
JC-DR-A-406.fm
7
COMPLETION
Carry out the following calibrations:
Calibrations - Checks
Job Card
Bias Calibration
Job Card CAL A003 - X-Ray Tube Focal Bias Voltage Scale Factor
Heating Current Calibration
Job Card CAL A001 - Heater Current Scale Factor
kV Scale Factor
Job Card CAL A011 - Calibration of kV Scale Factor
mAs X-ray Calibration
Job Card CAL A012 - X-Ray Tube Heater Current
AOP/SNR Check