Chapter 6
Page no. 392
JC-PHY-A-001.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A001 - Unpacking the Gantry and Generator Cabinet
6-2
Unpack the Generator Cabinet
1. Remove the securing straps from the generator cabinet.
2. For each of the two red clamping brackets securing the cabinet to
the pallet:
a. Remove the two bolts which hold the bracket to the pallet.
b. Remove the protective film from the strips of adhesive tape on
the bracket and on the bottom edge of the cover panel immedi-
ately above the panel.
c. Push the bracket firmly upwards as shown (C) so that it is held
up by the adhesive tape.
3. Remove the screws (B) which hold the two wooden blocks under-
neath the cabinet; remove the blocks.
4. Leave the two brackets in place (held to the covers by the adhesive
tape) until you are ready to remove the covers and connect the sys-
tem. Then remove them and place them with the other packing
materials for return.
5. Unpack the Gantry and remove the Gantry from the pallet (see
6. Carefully roll the cabinet (and the Gantry together) down the ramp and into the mammography room.
CAUTION
Take care that the generator cabinet does not catch and jump while moving down the
ramp. The center of gravity of the generator cabinet and the Gantry cabinet are high.
A
B
C