GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A009 - Opposite Side Rotative Arm Installation
Page no. 429
Chapter 6
JC-PHY-A-009.fm
•
Ground cable
29. Distribute evenly (36) the harness over the monitor cables on both sides of the monitor support.
30. Place the mouse at the back of the mouse support (so that there is enough cable for use of the
mouse on the entire mouse support).
31. Use pointed nose pliers to create an opening in the harness at 27 cm from the end of the harness.
35
36
36
27 cm