Chapter 6
Page no. 468
JC-PHY-A-015.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A015 - HT Cable Connection
4. Use a suitable pin wrench to tighten the cable securing nut (1) until resistance is felt. Try to move the
probe by hand. If any movement is possible, tighten the nut further. Continue until no movement is
detected.
5. The tank air vent screw (2) must be slightly open, to ensure that no excess pressure builds up during
long exposures. Tighten the screw by hand, then unscrew it to leave a 3 mm (1/8 inch) gap. Do not
remove the screw.