ERR SUB A003 - Gantry Error Messages
Page no. 907
Chapter 9
ERR-SUB-A-003.fm
152/009
DEBUG
ONLY
Internal 13.3V volt-
age is out of spec
limits
The internal 13.3V voltage is out of the specified lim-
its.
Ignore this error. If the error persists, contact your
OLC.
152/010
DEBUG
ONLY
Internal 5V DC volt-
age is out of spec
limits
The internal 5 V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
152/011
DEBUG
ONLY
Negative unregu-
lated voltage is out of
spec limits
The negative unregulated voltage (coming from
transformation) is out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
152/012
DEBUG
ONLY
Positive unregulated
voltage is out of spec
limits
The positive unregulated voltage (coming from trans-
formation) is out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
152/013
DEBUG
ONLY
Internal -5V DC volt-
age is out of spec
limits
The internal -5V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
152/014
DEBUG
ONLY
Internal 5V DC volt-
age is out of spec
limits
The internal 5 V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
152/015
DEBUG
ONLY
Internal reference
voltage for AD con-
versions is out of
spec limits
The internal reference voltage for A/D conversions is
out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
152/019
DEBUG
ONLY
Internal 0.6 volts is
bad. Measurement
circuit error
The internal 0.6 V is bad due to a measurement cir-
cuit error.
Ignore this error. If the error persists, contact your
OLC.
152/020
DEBUG
ONLY
Internal EEPROM
error
The PDU MCU has failed. This error is also gener-
ated when EEPROM is empty. This behavior is nor-
mal during the first 5 s after the PDU is first turned
on.
Ignore this error. If the error persists, contact your
OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card