Chapter 7
Page no. 558
JC-ELE-A-008.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
5. Once the transfer of the test patterns is complete, click the top right corner of the
Media Browser:CDR
window to return to the main Browser window. The Patient List in the Browser
now includes a Patient called
AAPM Test Patterns
, which contains the test patterns required to
check the luminosity of the 21-inch 3MP Monitor.
6. In the Browser, click the
CDR-read/write
icon and select
Detach
from the menu that appears. The
DVD-ROM drive of the ADS Computer opens.
7. Remove the Test Pattern CD-ROM and close the DVD-ROM drive.
8. From the Patient List in the Browser, select the
AAPM Test Patterns
Patient.
9. In the Study list, select the
1K-LUMIN
Study ID.
10. In the Series list, select Series
2
(Procedure Description
TG18-LN
).
AAPM Test Patterns
AAPM Test Patterns