Chapter 3
Page no. 48
Introduction.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Sub-Systems and Components
2-4-2.
Uninterruptible Power Supply (UPS)
To assure system safety in the event of disturbances in the mains power supply, the Senographe system
incorporates a UPS, housed in the Control Station cabinet.
During power fluctuations or brief interruptions, the UPS assures a continuous supply to the workstation
and the acquisition system. The UPS prevents mains disturbances from being transmitted to the system.
When a power failure occurs the UPS continues to supply power to the workstation until it has shut down
safely.
2-4-3.
External Connections
One Ethernet cable connection to the local hospital network is connected at the rear of the cabinet for
external communications (such as Insite connection):