Chapter 9
Page no. 1454
JC-DR-A-315.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A315 - International Keyboard
9. You are returned to the command prompt.
10. At command prompt, launch the Desktop Manager Configuration Tool again, by entering:
kdmconfig
The Desktop Manager Configuration Tool appears. Within the Desktop Manager Configuration Tool
you use the cursor keys to move up and down between the different options, and the space bar to
select an option. The currently selected option contains an
x
within the brackets.
11. At the first
kdmconfig
page, use the cursor keys to move to the
Xorg server
option, and press the
space bar to select that option. Press the
F2
key to continue.
12. At the first
Xorg Server Configuration Instruction
page that appears, press the
F2
key to continue.
13. You are returned to the command prompt.
14. At command prompt, ensure that the Control Station is in safe state for powering down, by entering:
shutdown -y -i 0
15. Power down the entire system, and proceed to
.
WARNING
Before proceeding to
, shut down correctly the equipment and switch off the
Senographe system hospital supply circuit breaker. Lock it in the off position or mark it
clearly with a warning sign to prevent others from applying power
.