GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A009 - Generator Configuration
Page no. 561
Chapter 7
JC-ELE-A-009.fm
Job Card ELE A009 - Generator Configuration
Chapter 7
1
INTRODUCTION
This module provides information for initial configuration of a Senographe system. It contains two sec-
tions:
•
Configuration procedures
. Procedures which are required to set up the system for local site condi-
tions.
•
Configuration check information
. For use only if you suspect that certain important settings have
been changed; most settings are made in the factory and do not need to be changed on site.
2
TOOLS
None
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 40 minutes
4
CONFIGURATION PROCEDURES
4-1
Prerequisites
The system must have powered up and reached a stable state with no checksum error displayed on the
X-ray Console.
•
If, on the first power up, CHECKSUM ERROR (GEN) appears, proceed as follows:
Reload the Generator saved parameters, using the delivered diskette (PN 2152250) or CD-R that
are to be used to save the generator parameters. This diskette is labeled with the same product loca-
tor number as your Senographe system. The CD-R is labeled as the Generator Backup.
The files are identified as Gxxxx.DAT (Generator), where xxxx is the factory serial number. There
should be no checksum error after a correct reloading of these parameters.
•
Perform a checksum (
SETUP/GENE/INSTAL/CKSUM/CKSUM
) and return to the Application mode.
4-2
Set Language
•
If necessary, change the X-ray Console display language by selecting:
SETUP/LANGUAGE/
Select
LANGUAGE,
where
LANGUAGE
is the desired language (FRANC, ENGL, DEUTS, ESPAN, ITALI,
PORTU, DANIS, FINNI, DUTCH, NORWI, or SWEDI ).
Note:
For compatibility with the English language documentation, you are recommended to select ENGL
during installation of the system; change to the local language when the installation is complete.