Chapter 9
Page no. 1044
JC-TSG-A-015.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A015 - Lift Movement Troubleshooting
2-2
Trouble Shooting Steering Guide For Type 2 Lift Screw Assembly
Use the following flow charts (illustration 4 and illustration 5) to isolate the issue causing the Lift move-
ment problems on systems with a Type 2 Lift Screw Assembly.
ILLUSTRATION 4 - LIFT TROUBLESHOOTING ENTRY - TYPE 2 LIFT SCREW ASSEMBLY
Does customer complain about
Error 102/024 or does Error 102/024 repeatly
appear in the error logs?
Perform the lift trouble shooting steps described
in illustration 5,
Lift Movement Problems Related to Errors 102/
013 and 102/014 - Type 2 Lift Screw Assembly
on
Are Errors 102/013 and 102/014 present on a
weekly basis?
Proceed to the Grease Checks in illustration 5,
and follow the flowchart as required.
OK
NOK
Lift Gas Spring Checks
See
NOK
Change Lift Gas Spring
See
.
NOK
Perform Lift Functional Checks.
See
,
. After the
checks, return back here.
OK
OK
OK
OK
OK
NOK
Perform Lift Functional Checks.
See
OK
Contact OLC.
No
Yes
Yes
No
Install/change Top Brake
(5198687) - See
A233 - Lift Screw Assembly Top
Requires the old system to have
2373567-8 Lift Board and Lift
CPLD version 10.0.1 or above
Perform Lift Functional Checks.
See
,
checks, return back here.
OK
Install Type 3 Lift Screw Assembly with Brake Motor (5337148)
See
Job Card D/R A232 - Lift Screw Assembly
Note:
The brake functionality requires the old system to have
2373567-8 Lift Board and Lift CPLD version 10.0.1 or above. Check
the Service Desktop home page and if the Lift Board CPLD version
is less than 10.0.1 you must also change the Lift Board according to