ERR SUB A003 - Gantry Error Messages
Page no. 865
Chapter 9
ERR-SUB-A-003.fm
3-2
Error Codes from 110/yyy: Rotation
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card
110/001
HALT
ROTATION RELAY
DEFECT
There is a relay defect at power-up. A relay (on the
rotation board) is mounted in parallel with the H-
bridge of the motor. This relay, when shut, creates a
short-cut of the H-bridge. At power-up, the node
does the following:
- injects 15 V to the H-Bridge
- shuts the relay and checks that there is 0 V
- re-opens the relay and checks that there is 15 V. If
not, this signal is produced.
1. Reboot the system. Check the error logs to see
whether the 110/001 ROTATION RELAY
DEFECT error occurs within the same time as
another POSITION ERROR (e.g. rotation or
compression).
2. If a POSITION ERROR (e.g. rotation or com-
pression) does exist within the same time, ignore
the 110/001 ROTATION RELAY DEFECT error.
Do not change the Rotation Board, and take
appropriate action to resolve the POSITION
ERROR.
3. If a POSITION ERROR (e.g. rotation or com-
pression) does not exist within the same time,
the 110/001ROTATION RELAY DEFECT error is
valid. Change the Rotation Board and then re-
calibrate the Rotation Arm.
D/R A221
CAL A021
110/006
HALT
ARM TURNED
RIGHT TOO FAR
The Arm is over the software specified limit for the
farmost right position. Occurs when the Arm position
is not between the software limits at power-up
(including a tolerance to be defined). Typically, this
can happen when Arm has been moved manually.
Shut down the Gantry, move lift down manually and
reboot the Gantry.
110/007
HALT
ARM TURNED LEFT
TOO FAR
The Arm is over the software specified limit for the
farmost left position. Occurs when the Arm position
is not between the software limits at power-up
(including a tolerance to be defined). Typically, this
can happen when Arm has been moved manually.
Shut down the Gantry, move lift up manually and
reboot the Gantry.