Chapter 9
Page no. 1382
JC-DR-A-287.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A287 - Collimator DC-DC Board
5. Power on the Gantry and check that no errors are raised on the Gantry LCD display.
6. Check that the LED states are as follows. For more information about the behavior and states of the
Collimator Control board LEDs, refer to
Collimator Control Board PL403
.
7. Reboot the Senographe system and verify that the Collimator blades move as expected.
7
COMPLETION
None.
LED Label
LED Color
LED Status
DS1
Green
ON
DS2
Green
OFF or ON
DS3
Green
ON
DS4
Green
ON
DS5
Green
ON
DS6
Green
OFF
DS7
Red
OFF
DS8
Red
OFF
DS9
Red
OFF
DS10
Red
OFF
CC1TP
TP20
TP17