Chapter 9
Page no. 848
ERR-SUB-A-001.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
094/003
E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on
Senographe which generates an error
on a signal/message expected from the
IDC task of the Senographe: Stop com-
mand at end of exposure not received
by the IDC task.
Check communication between IDC
and Senographe, perform comm
diagnostics, perform IDC diagnostics
094/004
E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on
Senographe which generates an error
on a signal/message expected from the
IDC task of the Senographe: XRayAu-
thorizedRT status not received by the
IDC task.
Check communication between IDC
and Senographe, perform comm
diagnostics, perform IDC diagnostics
094/005
E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on
Senographe which generates an error
on a signal/message expected from the
IDC task of the Senographe: Stop com-
mand not received by the IDC task at
the end of the Radiography.
Check communication between IDC
and Senographe, perform comm
diagnostics, perform IDC diagnostics
094/006
E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on
Senographe which generates an error
on a signal/message expected from the
IDC task of the Senographe: Prep com-
mand not received by the exposure
automaton (task T_APPLI)
Check communication between IDC
and Senographe, perform comm
diagnostics, perform IDC diagnostics.
If the problem persists, additionally try
adding DB9 Adaptors to the DB9
sockets on the IDC rear panel (see
Job Card TSG A012 - Use of DB9
Adaptors to Reduce Intermittent IDC
or ADS Communication Errors
094/007
E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on
Senographe which generates an error
on a signal/message expected from the
IDC task of the Senographe: XRayAu-
thorizedRT signal active in idle status
Check communication between IDC
and Senographe, perform comm
diagnostics, perform IDC diagnostics
096/019 I34 Thickness out of
AOP range
Thickness out of AOP range.
None (warning only)
096/020
I36 IFF threshold
out of range
IFF (IDC algorithm) threshold out of
range.
None (warning only)
096/021
I37 9 param model
deviate
Error on Scf factor for the pre-exposure Warning given in AOP mode between
pre-exp. and exposure, if the cor-
rected mechanical thickness is equal
to 0 or breast composition is out of the
range 0.6 to 1.4. Check consistency
between radiological and mechanical
thickness after pre-exposure. Check
gene. display.
096/023
I05 Exposure
aborted by mAs
level
mAs value performed for the pre-expo-
sure out of range
Reboot the full system.
Code
Message text
Description
Recommended actions