GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
Page no. 553
Chapter 7
JC-ELE-A-008.fm
7-2
Monitor Mode Configuration
The 21-inch 3 MP Monitor needs configuring to different settings to the factory defaults, so that:
•
Landscape mode is enabled (see
•
only the 1-DICOM mode is available and no other modes can be selected (see
and
)
•
the mode window only appears for 15 seconds (see
)
•
the luminosity of the monitor is checked to be 400 cd/m² (see
)
•
the control buttons are inactive (to prevent the Operator changing the settings) (see
)
Note:
If the 21-inch 3 MP Monitor control buttons are locked, you can unlock them according to the pro-
cedure described in
.
7-2-1
General Controls
This section descibes the general use of the 21-inch 3 MP Monitor control buttons and power LED. The
right hand side of the 21-inch 3 MP Monitor contains various buttons and a LED as shown below.
Control Panel
Existing
Mode
Setting
Item
Function
Description
1
Power Indicator LED
Indicates the monitor’s operational
status: Green - ON, Yellow - Power
Saving, not illuminated - OFF.
2
Power
button
Turns the monitor power on or off.
3
Navigational menu buttons
Up
,
Left
,
Right
, and
Down
buttons
for selecting different monitor configu-
ration menu items and adjustings
4
Enter
button
A single tap displays the ScreenMan-
ager menu, or confirms the selection
of a menu item.
A double tap saves adjusted values
and closes configuration menu.
5
Mode
button
Toggles between different preset dis-
play modes
Note:
The items in bold above are the names of the buttons used in the
procedures below within this Job Card.
1
2
3
4
5