MDS

 

 

Mobile Display System

 

 

 

 

Installation & Operator 

Manual 

Содержание MDS

Страница 1: ...Made in the USA MDS Mobile Display System Installation Operator Manual...

Страница 2: ...nd or international agencies regarding specific requirements and regulations applicable to the use of this equipment Image Diagnostics Inc certifies only the equipment After sale operating practices a...

Страница 3: ...ALLATION INSTRUCTIONS 6 6 1 Mount Two Main Monitors 7 6 2 Release Monitor Counterbalance 9 6 3 Shelf Adjustment 10 6 4 Clean Unit 11 6 5 Electronics Installation 11 7 OPERATOR INSTRUCTIONS 12 7 1 Movi...

Страница 4: ...eath personal injury or damage to the equipment Recyclable material Protective Ground This is the common tie point between the AC Electrical Power Cord Ground Frame Ground and Controller Ground SGS No...

Страница 5: ...tand its operation instructions emergency procedures capabilities including total range of motion and be aware of all potential safety hazards This manual should be accessible to all personnel install...

Страница 6: ...utlet not to exceed the rated amp values as stated on the MDS serial number label Safety Hazard Level Potential Consequences with Use DANGER Indicates an imminently hazardous situation which if not av...

Страница 7: ...operator inadvertently runs into the extended monitor arms Be aware of the arm s location at all times during the installation and while in use CAUTION Prior to each clinical placement confirm that t...

Страница 8: ...nternal components may result in increased emissions or decreased immunity of the equipment or system WARNING If MDS is used adjacent to or stacked with other medical electrical equipment the equipmen...

Страница 9: ...oks should ONLY be utilized during transportation of the unit Note The original factory balance weight is noted on the face of the VESA Monitor Support Bracket and the Face Balance Plates If Monitors...

Страница 10: ...s are secure and in place against the top and bottom Counterweight Bracket Flanges as shown in the photos below NOTE The upper and lower Locking Collars are used to secure the Arms for shipping only a...

Страница 11: ...6 With assistance from another individual lift the Monitor with the partially installed upper screws and slide the screws into the upper slots of the VESA Mounting Plates on the MDS arm 6 1 7 Apply 22...

Страница 12: ...d secure the counterweight Locking Collars are ready to be relocated from their storage positions 6 2 2 With a 3 16 Hex Key loosen the lower Locking Collars and lower them to the bottom of the slide r...

Страница 13: ...rvice for procedure steps NOTE There are inner and outer locking collars for each balance plate 6 3 Shelf Adjustment Tools 3 16 Hex Key a Level and an assistant Procedure 6 3 1 Remove all devices off...

Страница 14: ...ite generic household bleach in a solution of 5 25 sodium hypochlorite diluted between 1 10 and 1 100 with water o Alcohol generic o Envirocide Procedure 6 4 1 Unplug the MDS system prior to cleaning...

Страница 15: ...re pushed in and Magnets are engaged 7 1 4 Lift the arms so that the base of the monitor clears the upper most Device placed on the Top Shelf and rotate the face of one of the Monitors inward and gent...

Страница 16: ...r 7 2 2 Push and pull monitor to desired location within the limits of the articulating arms they are mounted to 7 2 3 To tilt monitor position downward hold the top of the monitor mount and push the...

Страница 17: ...manual and the proper tools Lubrication of this device is not required 8 1 Recommended Periodic Performance Checks Daily Inspect all external cables and electrical components for wear and damage Damag...

Страница 18: ...e appropriate lockout tag out procedure while performing any diagnostics and service on the unit 9 TROUBLESHOOTING Problem Symptom Possible Cause Remedy 1 MDS generates or receives electrical interfer...

Страница 19: ...t to 50 A Isolation transformer is certified to meet Medical Safety Standards ANSI AAMI ES60601 1 Environmental Operating Temperature Range 41 F to 95 F 5 C to 35 C Operating Humidity Range 10 to 85 r...

Страница 20: ...stics com 17 12 WARRANTY Warranty details for IDI Products can be obtained directly from Image Diagnostics Inc Image Diagnostics Inc 310 Authority Drive Fitchburg MA 01420 USA Or call IDI at 978 829 0...