Pressure Automated Calibration Equipment

Calibration manual - K0450 Revision B

PACE5000

 

PACE6000

 

PACE1000

 

Содержание PACE5000

Страница 1: ...GE Digital Solutions Pressure Automated Calibration Equipment Calibration manual K0450 Revision B PACE5000 PACE6000 PACE1000...

Страница 2: ...ons are subject to change without notice GE is a registered trademark of General Electric Company Other company or product names mentioned in this document may be trademarks or registered trademarks o...

Страница 3: ...ge Use suitably qualified calibration technicians and good engineering practice for all procedures in this publication Pressure Do not apply pressures greater than the maximum working pressure to the...

Страница 4: ...d airspeed i e That is IEC International Electrotechnical Commission IEEE 488 Institute of Electrical and Electronic Engineers standard 488 data in Inch kg Kilogram kts kn knot LCD Liquid crystal disp...

Страница 5: ...ets the requirements of all relevant European safety directives The equipment carries the CE mark This symbol on the equipment indicates that the user should read the user manual This symbol on the eq...

Страница 6: ...RE APPLYING PRESSURE EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT DO NOT EXCEED THE MAXIMUM WORKING PRESSURE O...

Страница 7: ...r the list of selectable pressure units Pressure units Factor hPa Pressure units Factor hPa mbar 1 0 cmH2O 20 C 0 978903642 bar 1000 0 mH2O 20 C 97 8903642 Pa N m2 0 01 kg m2 0 0980665 hPa 1 0 kg cm2...

Страница 8: ...es on Calibration 1 5 1 PACE Controller Connection 2 Gauge Reference 2 5 2 PACE Indicator Connection 3 Gauge Reference 3 6 Calibration Check All Ranges 4 6 1 Procedure 4 7 Calibration Adjustment 5 App...

Страница 9: ...rement uncertainty of the original calibration standard To preserve uncertainty of the PACE calibration checks and adjustments must be performed using a calibrator uncertainty of less than or equal to...

Страница 10: ...open to atmosphere Turn the valve to the closed position Gauge Reference If the pressure standard has a reference connection then connect this to the PACE reference port on the module manifold Otherw...

Страница 11: ...essure standard has a reference connection then connect this to the indicator reference port Otherwise the indicator reference port should be open to atmosphere Note Indicator reference port open to a...

Страница 12: ...e to the first pressure value and wait until this pressure as displayed on PACE is stable to less than 10ppm 0 001 Display filtering may be required 4 Compare the pressure value on the calibration sta...

Страница 13: ...e to be set on the pressure standard and to press OK when the applied pressure is stable Use the numeric keys to enter the precise applied pressure Note The display also shows throughout this procedur...

Страница 14: ...K0450 Revision B 6 EN English Appendix A PACE Menus and Screens A 1 PACE Controller Menus Calibration procedure next page...

Страница 15: ...PACE Series Calibration Manual EN English 7 K0450 Revision B A 2 PACE Controller Screens...

Страница 16: ...K0450 Revision B 8 EN English A 3 PACE Indicator Menus Calibration procedure next page...

Страница 17: ...PACE Series Calibration Manual EN English 9 K0450 Revision B A 4 PACE Indicator Screens...