PACE Series Calibration Manual
[EN] English
3
K0450 Revision B
5.2 PACE Indicator Connection
Connect the indicator for each pressure range as follows:
Connect the output of the pressure standard to the indicator port.
Gauge Reference
If the pressure standard has a reference connection, then connect this to the indicator
reference port. Otherwise the indicator reference port should be open to atmosphere.
Note:
Indicator reference port open to atmosphere or to reference port of the pressure
standard.
Note:
Optimum performance is achieved by connecting the PACE reference port to the
pressure standard.
SENSOR
PORT
PACE Indicator
Rear view of
instrument
Pressure Standard
REFERENCE
PORT