Performance Tester 

 

 

 

 
 
 
 
 
 
 

Stock No:   18708-2 

    June 2012 

 

Registered Community Design No: 001051940 

 

© Kane International Ltd 

Содержание 501C

Страница 1: ...KANE501C Performance Tester Stock No 18708 2 June 2012 Registered Community Design No 001051940 Kane International Ltd...

Страница 2: ...ING 5 1 6 POST TEST 5 1 7 OVERVIEW 6 2 MAINTENANCE 7 9 2 1 GENERAL MAINTENANCE 7 2 2 PERIODIC SERVICE 7 2 3 ANNUAL RE CALIBRATION 7 2 4 CLEANING 7 2 5 WATER TRAP 8 2 6 CHANGING PARTICLE FILTER 8 2 7 B...

Страница 3: ...due considerations of all the potential hazards 1 2 Pre test Checklist Clean particle filter Water trap and probe are empty of water All hose and thermocouple connections are properly secured Flue gas...

Страница 4: ...d then it will complete the zero process NOTE If you wish to use the same fuel as previously selected rotate the selector to Stand By at power on The fuel in use will be displayed and then blink If th...

Страница 5: ...nalyser and turn the analyser off and on again in fresh air 1 6 Post Test Remove the probe from the flue and allow the analyser to purge with fresh air until readings return to zero Be careful as the...

Страница 6: ...KANE501C Manual Page 6 1 7 Overview Fuel Test Type Indicator Water Trap LCD Display Selector Dial...

Страница 7: ...could result in physical degradation of the instrument This could alter the protection from personal injury this analyser provides to the operator Perform only those maintenance tasks that you are qu...

Страница 8: ...ng the Particle Filter This is a very important part of the analyser and should be changed regularly It prevents dust and dirt particles from entering the pump and sensors that will cause damage The f...

Страница 9: ...be appropriate to check the analyser prior to use The following procedure should be adopted Go through the normal start up sequence in the location where the equipment is to be used Switch on all loca...

Страница 10: ...ctors Meter does not respond to flue gas Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants Net temperature or Efficiency calculation incorrect Ambient tempe...

Страница 11: ...fuel type from the CO2 measurement It assumes that the combustion process is running fuel lean It is important to make other measurements to confirm that this assumption is always correct O2 Oxygen O...

Страница 12: ...KANE501C Manual Page 12 Fuels Available Selector Position Display Type L Oil LOIL Light Oil Propane PrOP Propane Nat Gas nGAS Natural Gas H Oil HOIL Heavy Oil Wood wOOd Wood Coal COAL Coal...

Страница 13: ...y 2 0 1 0 110 Excess Air 3 0 1 0 250 Pre Programmed Fuels Natural Gas Light Oil Propane Wood Heavy Oil Coal Dimensions Approximate Weight Handset Probe 0 5 kg 175 x 45 x 80mm 6mm diameter x 250mm long...

Страница 14: ...KANE501C Manual Page 14...