background image

A

UT

O

CL

A

VE

P

R

IM

A

QUICK GUIDE FOR ASSISTANT

Содержание PRIMA

Страница 1: ...AUTOCLAVE PRIMA QUICK GUIDE FOR ASSISTANT...

Страница 2: ......

Страница 3: ...RIMA ACCESSORIES I H x 1 x 2 N DESCRIPTION QUANTITY A RACK FOR TRAYS 1 PIECE B TRAYS 4 SET C DRAINAGE PIPE 2 PIECES D GASKET 1 PIECE E INSTRUMENT FOR TRAYS 1 PIECE F USER MANUAL 1 PIECE G DOOR KEY FOR...

Страница 4: ...AUTOCLAVE PRIMA OVERVIEW A Function panel B Clean water tank C Printer for paper report D Black door handle E Blue starting button F USB port for digital reports A B F C D E E 4...

Страница 5: ...PRIMA FIRST USE OF AUTOCLAVE 1 Open the door strongly pulling the black handle 2 Take out trays with accessories from the chamber 3 Remove protective plastic from trays and given accessories 1 2 3 5...

Страница 6: ...and warnings In case of anomalies the corresponding error code will be shown here See User manual appendix 2 page 42 D Lighting signals panel E sterilization programs F Graphics of sterilization cycl...

Страница 7: ...ing lighting signals Helix B D and Vacuum lighting signals This signal indicates that PRIMA is a B class autoclave This light is on when the clean water level in the tank is low This light is on when...

Страница 8: ...nd the indicator of class B Prima autoclave will be lightening accompanied by a short sound Moreover the indicator corresponding to the last selected program will lighten on the display I Open the doo...

Страница 9: ...if you want to set up the sterilization cycle you ll have to add distilled water before doing that STEP 1 Open the tank cover on the upper side of the sterilizer at right STEP 3 After pouring distille...

Страница 10: ...STEP 4 1 PUSH 2 PULL STEP 5 Water add by automatic suction has a length of 4 minutes press again START button if you want to interrupt automatic suction After adding water by automatic suction remove...

Страница 11: ...the water needs to be drained from the tank To drain dirty water from the tank do the following actions STEP 1 Take the autoclave water tube STEP 2 Open the door Connect the water tube to the Used wa...

Страница 12: ...ools are wrapped it s better to put them with paper side on the top and don t overload the trays 3 Put the trays in the chamber pushing them in the removal guides and verify the tray is straight Put t...

Страница 13: ...10kPa sterilization length 5 mins drying time 9 mins C Lighting signal indicating prion sterilization program 134 C 18 mins for prions and virus vacuum in 3 times temperature 134 C pressure 210kPa ste...

Страница 14: ...related signal brightens To change steri lization program go on pressing PROG button until the chosen signal program brightens 2 Once the sterilization program has been selected press START button to...

Страница 15: ...end of sterilization cycle the writing PASS will light up on the display If not the writing Er will light up In this case please read the user manual page 38 appendix 4 2 Please wait few minutes and o...

Страница 16: ...the printer B Through DIGITAL REPORT Remove the USB from its port and connect the USB in the pc so you can download the digital report If the writing ERROR Code displays on the screen so it means the...

Страница 17: ...clean pipes valves and steam generator The time of the cycle is 1 minute Pipeline cleaning lighting signal To start the pipeline cleaning program press TEST button 1 until you select the pipeline cle...

Страница 18: ...he autoclave Make sure to leave the door partially open to let the machine cool If you want to set a new sterilization cycle do once again the procedure previously described NOTE If standby mood or en...

Страница 19: ...asive cleaners on the gasket or mating surface of the chamber B At least once a week clean chamber trays and rack with a clean damp cloth C Clean the mating surface of the chamber with clean damp clot...

Страница 20: ...PRIMA Via Umbria 10 20090 Segrate Milan ITALY info galbiati com www galbiati com...

Отзывы: