background image

ModeL:FL-100 
 

Pulse Oximeter 

 

General Description   

 
Oxygen  Saturation  is  a  percentage  of  Oxyhemoglobin  (HbO

2

)  capacity, 

compounded with oxygen, by all combinative hemoglobin (Hb) capacity in 
blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a 
very  important  parameter  for  the  Respiratory  Circulation  System.  Many 
respiratory  diseases  can  result  in  oxygen  saturation  being  lowered  in 
human  blood.  Additionally,  the  following  factors  can  reduce  oxygen 
saturation:  Automatic  regulation  of  organ  dysfunction  caused  by 
Anesthesia,  Intensive  Postoperative  Trauma,  injuries  caused  by  some 
medical  examinations.  That  situation  might  result  in  light-headedness, 
asthenia, and vomiting. Therefore, it is very important to know the oxygen 
saturation of a user so that doctors can find problems in a timely manner. 
The fingertip pulse Oximeter features small size, low power consumption, 
convenient operation and portability. It is only necessary for a user to put 
one of his fingers into the fingertip photoelectric sensor for diagnosis, and a 
display screen  will  show  oxygen  saturation.  It  has  been  proven  in clinical 
experiments that it also features high precision and repeatability. 

Measurement principle 

Principle of the Oximeter is as follows: A mathematical formula is 
established making use of Lambert Beer Law according to Spectrum 
Absorption Characteristics of Reductive hemoglobin(RHb) and 
Oxyhemoglobin (HbO

2

) in glow and near-infrared zones. Operation 

principle of the instrument: Photoelectric Oxyhemoglobin Inspection 
Technology is adopted in accordance with Capacity Pulse Scanning and 
Recording Technology, so that two beams of different wavelength of lights 
(660nm glow and 940nm near infrared light) can be focused onto a human 
nail tip through a clamping finger-type sensor. A measured signal obtained 

by a photosensitive element, will be shown on the Oximeter’s display 
through process in electronic circuits and microprocessor shown on the 

Oximeter’s display through electronic circuits and a microprocessor. 

Diagram of Operation Principle 

1. Red and Infrared-ray Emission Tube 
2. Red and Infrared-ray   
3. Receipt Tube 

Precautions for use 

Do not use the pulse oximeter in an MRI or CT environment 

Do not use the pulse oximeter in situations where alarms are required. 

The device has no alarms. 

3  Explosion hazard:

 Do not use the pulse oximeter in an explosive 

atmosphere. 

The pulse oximeter is intended only as an adjunct in user assessment. It 

must be used in conjunction with other methods of assessing clinical 
signs and symptoms. 

Check the pulse oximeter sensor application site frequently to determine 

the positioning of the sensor and circulation and skin sensitivity of the 
user.   

Do not stretch the adhesive tape while applying the pulse oximeter 

sensor. This may cause inaccurate readings or skin blisters. 

Before use, carefully read the manual. 

The pulse oximeter has no SPO

2

 alarms; it is not for continuous 

monitoring.   

Prolonged use or the user’s condition may require changing the sensor 

site periodically. Change sensor site and check skin integrity, circulatory 
status, and correct alignment at least every 4 hours. 

10 

Inaccurate measurements may be caused by autoclaving, ethylene 

oxide sterilizing, or immersing the sensors in liquid. 

11 

Significant levels of dysfunctional hemoglobins (such as carbonxy- 

hemoglobin or methemoglobin) may cause inaccurate readings. 

12 

Intravascular dyes such as indocyanine green or methylene blue. 

13 

SPO

2

 measurements may be adversely affected in the presence of high 

ambient light. Shield the sensor area (with a surgical towel, or direct 
sunlight, for example) if necessary. 

14 

Excessive user movement may cause inaccurate readings. 

15 

Venous pulsations may cause inaccurate readings. 

16 

High-frequency electrosurgical interference may cause inaccurate 

readings. 

17 

Placement of a sensor on an extremity with a blood pressure cuff, 

arterial catheter, or intravascular line. 

18 

The user has hypotension, severe vasoconstriction, severe anemia, or 

hypothermia. 

19 

The user is in cardiac arrest or is in shock. 

20 

Fingernail polish or false fingernails may cause inaccurate SPO

2

 

readings. 

Follow local ordinances and recycling instructions regarding 
disposal or recycling of the device and device components, including 
batteries. 

Product Properties 

1  Operation of the product is simple and convenient 

2  The product is small in volume, light in weight and convenient in 

carrying. 

3  Power consumption of the product is low and the two AAA batteries can 

be operated continuously for 24 hours.   

4  A low voltage warning will be indicated in visual window when battery 

voltage is so low that normal operation of the oximeter might be 
influenced. 

5  The product will automatically be powered off when no signal is in the 

product for longer than 16 seconds. 

Product Operation Scope 

 
Fingertip PULSE OXIMETER is a portable non-invasive, spot-check, 
oxygen saturation of arterial hemoglobin (SPO

2

) and pulse rate of adult 

and pediatric user at home, and hospital (including clinical use in 
internist/surgery, Anesthesia, intensive care etc). It is not for continuously 
monitoring. 
 

Operation Instructions 

1  Install two AAA batteries into battery compartment correctly. 

2  Place clamp over finger nail as the following diagram. 
3  Insert one finger into rubber hole of the Oximeter fully. 

4  Press the switch once on the front panel. 

5  Finger and body should not tremble during measuring. 

6  Read correct data from display screen. 

After turn on the oximeter, each time you press the function button, the 
oximeter will switch to another display mode, there are 4 display modes 
totally.   

Long press the function button to enter the settings screen. 

 

Short press to switch among the settings, long press to change the 
current settings of sounds and alarm limits. 

Notes:   

 When your finger    insert into the Oximeter, your nail surface must be 

upward. 

 The results may be wrong if you did not plug the finger thoroughly in the 

Oximeter. 

 Please use medical alcohol to clean the rubber touching the finger inside 

of Oximeter, and clean the test finger using alcohol before and after each 
test. (The rubber inside of the Oximeter belongs medical rubber, which 
has no toxin and no harmful to the skin). 

 

Brief Description of Front Panel 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

The Pulse bar graph displays corrernding with the user’s pulse beat. The 

height of the bar graph shows the user’s pulse strength.   
 

Product Accessories 

1. 

Lanyard 

2. 

Two batteries 

3. 

One user manual

 

Battery Indicator 

Functional oxygen 
saturation of arterial 
hemoglobin 

Bargraph

 

Pulse Rate

 

Plethysaogram

 

Power On

 

Отзывы: