5.5 Syringe Replacement
Items to be prepared:
Hypodermic syringe
A tube between a syringe and a syringe valve
Syringe, Beaker, Distilled water, Wash fluid
Syringe (175 µL) (P/N 893-0815)
Syringe (70 µL) (P/N 893-0817)(option)
(1) Remove the sealing screw connected to the port 4 of the
syringe valve.
Plunger setscrews
Syringe valve
Sealing screw
Hook
Syringe
Syringe mount
Fig. 5-6
Syringe Valve
5 - 8