5 - 25
(4) Put seal guide pins into holes of the auto purge valve seal,
and install a new auto purge valve seal.
NOTE
: Avoid touching the rotor seal and the stator directly with
your hands. It may influence the result of analysis.
(5) Put stator guide pins into holes of the stator, and tighten lock
nuts with the hexagon wrench. Tighten alternately lock
nuts (3 pcs) little by little.
NOTE
: Tighten 3 fixation screws equally to prevent failure
operation of a valve, wear of a rotor seal and damage.
(6) Connect each tube to the auto purge valve in the reverse
order of their removal.
(7) Start delivering liquid of the pump via the auto purge valve.
(8) Make sure that there is no leakage after the seal replacing.
NOTE
: Avoid moving a valve when the inside of the valve is
dried. A seal wears away abnormally.
NOTE
: Decrease of liquid in the valve leads lack of lubrication at
seal surface. And normal movement is disturbed.
NEW auto purge valve rotor seal
Fixation screw
NEW auto purge valve stator
Auto purge valve main unit
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...