5.2.1
(4) Operation
Check
Plug the power cord into the instrument and turn on the
power switch. Make sure the lamp comes on.
NOTE
: The
D
2
lamp does not come on if it has been
mounted improperly.
(5) On the monitor screen 2, check the wavelength (WL) and
reference energy (REF), and write down their values as a
record of the D
2
lamp after use. This record can be used
as a reference for the next D
2
lamp replacement and as a
historical record of D
2
lamp replacement.
Monitor Screen 2
250 191000 202000 ON
WL
SAMP
REF
CELL
[I]
(6) Enter the lamp replacement date as follows.
(a) Press key, and the following screen will appear.
GLP
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
GLP (LOGBOOK=1, KEY LOCK=2,
(b) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
1
LAMP SELECT (D2=1, W=2, Hg=3)
(c) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
1
LOGBOOK (REPORT=1, RESET=2)
(d) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
<1–12> _9
15
2010
D2 LAMP CHANGE (MM DD YYYY)
5 - 7
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...