Create a time program as indicated below for measuring the
Raman spectrum of water.
TIME
Ex
Em
BASE (A/Z=1, HOLD=2)
0.0 350
450
1
0.5
360
0.6
370
0.7
375
0.8
380
0.9
385
1.0
390
1.1
395
1.2
400
1.3
405
1.4
410
1.5
415
1.6
430
1.8
450
2.0
NOTE
: The above time program shows an
example where the minimum FL value
was obtained at 380 nm and the
maximum at 400 nm. For the top Em
of the program, set a wavelength that
includes the wavelengths where the
maximum and minimum FL values were
obtained. (For instance, if the maximum
FL value was obtained at a wavelength
of 397 nm, then the wavelength at 1.1
minutes should be set at 397 nm.)
4 - 13
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...