5.2.2
(c) Remove the Hg lamp.
Hg lamp
(d) Fit a new Hg lamp and insert it into the instrument.
(e) Attach the Hg lamp cover at the original position.
In succession, attach the lamp cover at the original
position.
(4) Check the wavelength accuracy of Hg lamp, referring to
4.4.1.
(5) Enter the lamp replacement date as follows.
(a) Press key, and the following screen will appear.
GLP
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
(b) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
LAMP SELECT (D2=1, W=2
,
Hg=3)
1
(c) Press [numerical] key [3], and then the [ENT] key.
The following screen appears.
LOGBOOK (REPORT=1, RESET=2)
1
(d) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
Hg LAMP CHANGE (MM DD YYYY)
<1–12> _9
15
2010
5 - 13
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...