Lamp protrusion
(faces instrument front panel)
Metal wire
Fig. 5-7 Orientation of Lamp Bulb Protrusion
(i) As shown in Fig. 5-6, put the anode terminal (marked
“+”) of the lamp into the metal fixture on the ceramic
plate. Then, using the anode nut, mount the lamp so
that the protrusion on the bulb is positioned as shown
in Fig. 5-7. Refer to Fig. 5-4 for a view of the lamp
when assembled.
NOTE
: Mount the xenon lamp so that the protrusion on
the bulb faces the instrument front panel (Fig. 5-4).
A mistake in the orientation may cause an
increased light loss and preclude obtaining the
desired performance from the instrument.
(j) After mounting the Xe lamp in the lamp house,
assemble it into the main body (Fig. 5-2).
Raise the lamp house and insert it into the main body,
then lower it and press it downward into place.
The lamp house will not move if installed correctly.
Finally, secure the lamp house with retaining screws
and connect its connector with the main body.
5 - 11
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...