5 - 5
5.3 Replacement of Pump Seal
<Required articles>
Pump seal
(P/N: 891-1213x 2)
Hexagon wrench (P/N: N315005, Size: B4)
Workshop tools (P/N: 810-1353)
Receiving tray
Remover N
(P/N: 893-1372)
Wash bottle
Gauze or soft cloth
Ultrasonic cleaner
(1) Disconnect the inlet tube and outlet pipes (1)(2) from the
pump heads using a 1/4 inch double-end wrench of
workshop tools.
(2) Loosen the hexagon screws alternately with a hex wrench
and remove them.
(3) Carefully pull out the pump head toward you with both
hands.
NOTE
:
Take care not to damage the plunger.
Pump head
Outlet pipe
(1)
Outlet pipe (2)
Pump head
Spring washer
Hexagon screws
Hex wrench
Plunger wash mechanism
Guide
Wash mechanism
plunger guide
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...