3 - 63
3.5.2 LCD Contrast Setting
For LCD contrast setting on the UI pad, take the procedure
shown below.
SETUP: PUMP MODE CONTRAST S.NO
1 FLOW 4 02AB-111
SETUP ITEMS (PUMP=1 CONTRAST=2 D.MIX=3)
2
ENT
LCD CONTRAST (1 - 7)
4
Select CONTRAST
2 ENT
SET UP
UTLTY
POWER ON +
Enter contrast intensity
4 ENT
Press the POWER and SET UP UTLTY keys simultaneously.
LCD contrast setting screen will be displayed.
Continue pressing SET UP UTLTY until the setting screen appears.
POWER OFF
POWER ON
SETUP: PUMP MODE CONTRAST S.NO
1 FLOW 4 02AB-111
Instrument starts with the set contents.