2. Place for Installation
For installation of the instrument, select a place which meets the
following conditions.
(1) Operating
temperature:
Within 4 to 35
°
C and with minimum temperature variation
during measurement. Utmost care should be used for
preventing condensation.
(2) Operating humidity: 25 to 85%
(3) Atmospheric
gas
(a) Adequate ventilation is required.
(b) The room shall not be filled with highly corrosive gas of
acid or alkali.
(c) The room shall not be filled with gas of an organic
solvent (particularly, benzine and paint thinner) which
dissolves coating.
(4) Other general precautions on instrument
(a) Shall not be exposed to direct sunlight.
Otherwise, performance degradation or coating
discoloration might be caused. Avoid a place near
windows as far as possible.
(b) Shall not be exposed to draft.
(c) Shall be free from strong vibration or shock which can
be felt by your body.
(d) Shall be free from a heating device such as gas burner,
electric heater or oven in the vicinity.
(e) Shall be away from strong electric field generating
equipment (such as electric welder, high frequency
electric furnace and pole transformer).
APPENDIX 1 - 2
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...