3.4.1
Next, set the maximum wavelength of scan range.
Press [numerical] keys (within [2][5][0] to [9][0][0]), and then
the [ENT] key.
* Enterable range: Minimum scan range is 50 nm.
Measuring span is max. 400 nm.
Input the number of spectral data.
Press [numerical] key (within [1] to [3]), and then the [ENT]
key.
During spectrum acquisition, the following screen is
presented.
Upon completion of spectral data storage, the monitor
screen automatically appears.
NOTE
: The stored spectral data is a value after automatic
subtraction of the background spectrum.
Therefore, before spectrum measurement, correct
background spectrum should be stored in
memory. If acquisition of spectral data has been
attempted without acquisition of background
spectrum, the following screen appears.
<250 – 600>
200 – 600
MEMORIZE SPECTRUM NO. (1–3)
_
SCANNING SPECTRUM NO.: 1 MODE: ABS
WL: *** nm
COLLECT “BACK GROUND” SPECTRUM FIRST
PRESS “CL” KEY TO CLEAR MESSAGE
SCANNING WL RANGE (WL1 WL2) nm ABS
3 - 31
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...