(2) Aspiration of wash fluid1
The syringe aspirates [vd :Dead volume (inside parameter)] of
fluid1. Switch the syringe valve to the wash port side (port1)
after aspiration of the wash fluid1.
When [wash before sample aspiration] is set, refer to the
following operation:
Washing of the needle outer wall and the injection port flow path.
Deliver the wash fluid1 with the wash pump, move the needle
into the wash port, and wash the needle outer wall.
Wash the injection port flow path with the wash fluid1 after the
needle has moved into the wash port (the same as 2.3.3 Cut
injection method procedure (9)).
And when [fluid2 wash] is set, wash the injection port flow
paths using the wash fluid2 (the same as 2.3.3 Cut injection
method procedure (10)).
Buffer tube
Injection valve
vd :Dead volume (the instrument inside parameter)
vi :Injection volume (the sample net volume to the flow path)
vw :Waste volume (the air volume of the lead and rear part of a sample)
va :Air volume (the air volume which sample is sandwiched.)
vd
Syringe valve
Syringe
5
1
3
2
4
Plunger
wash
Needle
Injection port
5
4
3
2
1
6
Waste
solution
Loop
Waste
solution
Wash port
Sample
Rack
P
Wash
fluid2
Wash
fluid1
From Pump To Column
1
3-way solenoid valve
D
2 - 37
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...